Chiptech
Chiptech
S72400230 Demo SEVEN Base Unit · Deployed
← Registry Deployed Synced NZ · NZD

S72400230

SEVEN Base Unit · hardware revision C · built 13 Jun 2024 · managed by Tākiri Health Trust

Recall RC-26-0002 SEVEN rev C — backup battery retention Identified

Rev-C charge termination reduces backup capacity to about 11 hours against a rated 40. Affected units are to be swapped on a scheduled visit, not left in service.

Identity
Everything printed on the label and everything SmartCare holds.
Serial S72400230
Product SEVEN Base Unit (SEVEN-BU)
Hardware revision C
IMEI 3500 0004 2113 957
CSID / SmartCare external id CS715702
ICCID 8964 0010 4390 36
IMSI 5304273342560
Phone number +64278799441
SIM provider One NZ
Network One NZ
Firmware 7.4.2
Configuration CFG-TAK-v4
Voice file EN-NZ-3.0
Ownership & position
Managing organisation Tākiri Health Trust
Commercial owner Tākiri Health Trust
SmartCare organisation ORG-TAK-001
Location In the field
Last heard from 14 h ago · 22 Aug 2026, 1:57 pm
Warranty In warranty until 13 Jun 2027 · 294 days left
Service
Billing →
This device is live with nothing billing against it.

It is in the field and reporting, and no monthly charge has ever been raised for it. This is exactly what the billing reconciliation looks for.

Where it came from
Built 13 Jun 2024
Production order Before this system
Shipped on
Activated 29 Jun 2024
SmartCare sync
Queue →
No synchronisation attempts recorded.
Service history
0 RMAs
This device has never come back.