What was found

SEVEN rev-C base units built between 15 January and 20 December 2024 lose backup capacity to roughly 11 hours after two years in service. Traced to a charge-termination fault in the rev-C charging circuit.

Containment and corrective action

Rev D corrects the charge termination in hardware. Firmware 7.4.2 adds a capacity self-test and reports it to SmartCare. Recall RC-26-0002 raised against the affected build window.

Recall campaigns raised from this
CampaignProductBuild windowAffectedActionedStatus
RC-26-0002
SEVEN rev C — backup battery retention
SEVEN Base Unit 15 Jan 2024 – 20 Dec 2024 40 23 Active